• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW MEDICAL LTD. PICO SINGLE USE NPWT 10X20CM; NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW MEDICAL LTD. PICO SINGLE USE NPWT 10X20CM; NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP Back to Search Results
Catalog Number 66800951
Device Problem Loose or Intermittent Connection (1371)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/16/2019
Event Type  malfunction  
Event Description
It was reported that that the nurse went to assess 2 separate patients in the tcu and the port had been completely detached from the dressing, leaving a hole in the dressing where the port would normally be anchored.The problem was solved by removing the dressings and the impact caused was that the patients didn¿t receive the negative pressure as hoped for.
 
Manufacturer Narrative
H2,h6: we have now concluded our investigation into this complaint.No valid lot number was provided, therefore a review could not be carried out to confirm that the product had been manufactured in accordance with defined procedures and specifications.No samples were returned for analysis, however visual inspection of the provided photograph confirms that the soft port has come away from the pico dressing.Without return of the sample relating to the complaint, a root cause analysis cannot be carried out to find a potential cause for this issue.During the manufacturing process, regular tests are used to ensure that the machine which sticks the soft port to the dressing is working as it should.These tests include pull tests, where a weight is attached to the end of the soft port to make sure that the strength of the bond is up to specifications, and a leak test, to check that the adhesive which hold the two components together doesn¿t have any gaps in it.If these tests don¿t meet specifications, the product is not released to market.Smith and nephew acknowledge customer concern and are continually investigating ways to develop and improve our products.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PICO SINGLE USE NPWT 10X20CM
Type of Device
NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull
UK 
MDR Report Key8681379
MDR Text Key147455334
Report Number8043484-2019-00389
Device Sequence Number1
Product Code OMP
Combination Product (y/n)N
PMA/PMN Number
K111170
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 07/22/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/09/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number66800951
Date Manufacturer Received07/18/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-