Catalog Number 190618 |
Device Problem
Melted (1385)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 05/20/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
|
|
Event Description
|
A fresenius (b)(4) technician reported that a fresenius 2008k2 hemodialysis (hd) machine had a melted component.A patient was not connected to the machine at the time of the incident and there was no harm to any patients or individuals because of this malfunction.Additional information was requested however to date has not been provided.
|
|
Event Description
|
Additional information was received by email response from the fresenius mexico technician.The machine was reviewed and repair was initially pending due to lack of spare parts.The power supply was replaced and functional tests were passed.The machine went through disinfect and was ready for use.No parts were available to be returned to the manufacturer for evaluation.
|
|
Manufacturer Narrative
|
Plant investigation: no parts were returned to the manufacturer for physical evaluation.A fresenius mexico technician performed an on-site investigation and replaced the damaged power supply to repair the machine.The machine passed functional testing, was disinfected and has been returned to service.A records review was performed on the reported serial number.An investigation of the device manufacturing records was conducted by the manufacturer.There were no non-conformances, or any associated rework identified during the manufacturing process which could be related to the reported event.In addition, the device history record (dhr) review confirmed the results of the in-progress and final quality control (qc) testing met all requirements.
|
|
Search Alerts/Recalls
|