• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS, INC. 1818910 MBT REV TIBIAL VIEW PLT SZ3; KNEE INSTRUMENT : SIZING/MEASURING INSTRUMENTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS, INC. 1818910 MBT REV TIBIAL VIEW PLT SZ3; KNEE INSTRUMENT : SIZING/MEASURING INSTRUMENTS Back to Search Results
Catalog Number 217865130
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/22/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Tibial view plate has snapped in half.Please see.Information is that this item has only been used a handful of times.This case has been reported by the loan kit technician during loan kit inspection.It has been forwarded to depuy engineering for assessment.No further information can be obtained as the case was not reported by the customer.Customer reference/po/service (b)(4), patient status/ outcome / consequences: unknown, was other medical intervention (e.G.X-rays, additional procedures, prescriptions, otc, revision) required: unknown, is the patient part of a clinical study: unknown, (b)(4).Device property of:none, device in possession of: none.By checking this box i certify that all information that are known/available has been disclosed.If any new information will be made available, the additional information will be submitted through cst.True.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Examination of the returned view plate confirmed the reported event the device has broken into two pieces.Root cause attributed to the device being worn from normal use and servicing.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : null.Device history batch : null.Device history review : null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MBT REV TIBIAL VIEW PLT SZ3
Common Device Name
KNEE INSTRUMENT : SIZING/MEASURING INSTRUMENTS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic drive
warsaw IN 46582 0988
MDR Report Key8684469
Report Number1818910-2019-95627
Device Sequence Number2268912
Product Code LXH
UDI-Device Identifier10603295096672
UDI-Public10603295096672
Combination Product (Y/N)N
Initial Reporter CountryAS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/22/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number217865130
Device Lot NumberSO2034337
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/07/2019
Initial Date Received by Manufacturer 05/22/2019
Supplement Date Received by Manufacturer08/05/2019
Initial Report FDA Received Date06/10/2019
Supplement Report FDA Received Date08/14/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-