| Model Number |
9733857 |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
08/01/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Other relevant device(s) are: product id: 9733597, serial/lot #: (b)(4).Analysis of the 9733597 found the connection cabling/hardware was damaged.The cable jacket was damaged near the lemo connector exposing a small amount of shield wire but no bare conductors.Otherwise, the cable passed a continuity test with no opens or shorts detected.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information regarding a navigation system while outside of procedure.It was reported that the axiem cable cover was found broken during inspection.There was no patient present when the issue occurred.
|
| |
|