| Model Number |
ZCT300 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Age/date of birth: unknown.Sex/gender: unknown.Date of event: unknown, not provided; best estimate is between (b)(6) 2019 and (b)(6) 2019.(b)(4).All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that a model zct300 23.0 diopter intraocular lens (iol) was implanted in the patient's operative eye on (b)(6) 2019.It was later explanted on (b)(6) 2019 due to myopic outcome.At explant the incision was enlarged, and the explanted lens was exchanged for another zct300 lens but a lower diopter (21.0).Reportedly, the patient is doing good post operatively.No further information provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the product was not returned to the manufacturing site; therefore, the complaint issue reported was not verified.Manufacturing records review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specification.Labeling review: the directions for use (dfu) were reviewed.The dfu adequately provides instructions, precautions, and warnings for the proper use and handling of the product.Historical data analysis: a search of complaints revealed no similar complaints were received for this production order number.Conclusion: as a result of the investigation there is no indication of a quality product deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
|
| |
|