| Model Number |
MMT-523NAP |
| Medical Device Problem Codes |
Device Difficult to Program or Calibrate (1496); Date/Time-Related Software Problem (2582)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/05/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the buttons were not responding.Customer¿s blood glucose was 210 mg/dl at the time of incident.Customer stated that the time was not advancing.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device received with intermittent button response due to corroded keypad traces.No button error alarm during testing.Device passed self test.No frozen screen noted during test and time clock advances properly.
|
| |
|