Device passed the rewind test, prime test and excessive no delivery test.The stop (idle) current and run current measurement tests are within specification.Device also passed self test, off no power alarm test and a21 error test.Unable to perform the displacement test, basic occlusion test, occlusion test and delivery accuracy test due to completely broken off reservoir tube lip.Unable to verify the no delivery alarm during the basic occlusion test and on the occlusion test due to completely broken off reservoir tube lip.No unexpected no delivery alarm note during the excessive no delivery test.The test p-cap and reservoir will not lock in place in the reservoir compartment due to completely broken off reservoir tube lip.Device also had a missing reservoir tube o ring, minor scratched display window, cracked battery tube threads, cracked case at the reservoir tube window corners and cracked reservoir tube window.Device uploaded properly using carelink.
|