• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER GMBH BIOLOX OPTION, HEAD, XL, 32/+7, TAPER 12/14; BIOLOX OPTION HEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER GMBH BIOLOX OPTION, HEAD, XL, 32/+7, TAPER 12/14; BIOLOX OPTION HEAD Back to Search Results
Model Number N/A
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Information (3190)
Date of Event 05/03/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Concomitant medical products: oblique liner 32 mm i.D.Size gg for use with 48 mm o.D.Size gg shell; catalogue: 00875500832 and lot#63635189, therapy date: (b)(6) 2018.The manufacturer did not receive x-rays, or other source documents for review.The manufacturer did not receive the device for investigation.The device history records were reviewed and found to be conforming.Additional information has been requested and is currently not available.A cause for this specific event cannot be ascertained from the information provided.As soon as supplemental information becomes available an updated report will be submitted.  (b)(4).
 
Event or Problem Description
Patient was implanted on the right side and underwent reduction due to dislocation.
 
Event or Problem Description
No change to previously reported event.
 
Additional Manufacturer Narrative
Device history record (dhr) was reviewed and no discrepancies were found.No trigger considering the following event is identified: dislocation.Event summary: it was reported that product was implanted on (b)(6) 2018 and underwent closed reduction due to 1st dislocation on (b)(6) 2018.Review of received data medical records were reviewed by the health care professional.The review identified the following: primary op notes dated (b)(6) 2011 were reviewed and no complications were noted.Revision op notes dated (b)(6) 2018 were reviewed and identified patient was revised due to failed mom construct.Medical notes dated (b)(6) 2018 were reviewed and identified patient experienced a dislocation of the right hip prosthesis requiring a closed reduction under conscious sedation.Medical notes dated (b)(6) 2018 were reviewed and identified patient experienced a second dislocation of the right hip prosthesis.A total right hip prosthesis is dislocated with the femoral component displaced superiorly.This is most likely an anterior dislocation.Devices analysis no product was returned to zimmer biomet for in-depth analysis.Review of product documentation" this device is intended for treatment.Devices analysis" no product was returned to zimmer biomet for in-depth analysis.Conclusion summary: it was reported that product was implanted on (b)(6) 2018 and underwent closed reduction due to 1st dislocation on (b)(6) 2018.In vivo time of the device is 2 months.Patient experienced a second dislocation on (b)(6) 2018.The quality records show that all specified characteristics (material, dimensions, surface, etc.) have met the specifications valid at the time of production.The investigation results did not identify a non-conformance or a complaint out of box (coob).Neither x-rays nor devices or photos of the explanted implants were received; therefore the condition of the components is unknown.Therefore, based on the given information and the results of the investigation, we were not able to identify a specific root cause for this issue.The need for corrective measures is not indicated and zimmer gmbh considers this case as closed.Zimmer biomet¿s reference number of this file is (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOLOX OPTION, HEAD, XL, 32/+7, TAPER 12/14
Common Device Name
BIOLOX OPTION HEAD
Manufacturer (Section D)
ZIMMER GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ  8404
MDR Report Key8685950
Report Number0009613350-2019-00373
Device Sequence Number2511130
Product Code LZO
UDI-Device Identifier00889024430501
UDI-Public00889024430501
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K073567
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberN/A
Device Catalogue Number00-8777-032-04
Device Lot Number2927640
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 05/19/2019
Supplement Date Received by Manufacturer11/19/2019
Initial Report FDA Received Date06/11/2019
Supplement Report FDA Received Date11/20/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Weight68
-
-