| Model Number |
866030 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/23/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A follow up report will be submitted once the investigation is complete.
|
| |
|
Event or Problem Description
|
|
The customer reported that room 306 alerted a vtach on (b)(6) 2019 at 9:47:03.There was a delay of alarms via the paging system.No information was received indicating the device was not a contributing factor in the event.The type of adverse event was not provided at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
A philips product support engineer (pse) obtained the paging system (em) logs.The pse reviewed the logs and concluded that the system functioned as designed.The investigation found that one alarm messages was delivered from bed 306 wmc to 23 may 2019 at 09:47:20am est (iem system time) to the zetron paging terminal for pager pin¿s = 2500, 1665, 990, 9998: an examination of he overall iem system found that the computer cpu is only one core, while the minimum specifications require two cpu cores.There was no product malfunction; this is considered a user issue.The system had sent a page for the alarm on 23 may 2019 at 9:47:20am for the vtach alarm occurring at 9:47:03am on 23 may 2019.The pse also emailed the customer requesting for the iem server to have at least one more cpu core added to the computer.
|
| |
|