| Catalog Number |
504-02-62E-L |
| Medical Device Problem Codes |
Material Deformation (2976); Packaging Problem (3007)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/13/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Review of the device history records indicate that devices were manufactured and accepted into final stock with no reported discrepancies.There have been no other events for the lot referenced.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.Device evaluated by manufacturer? not available.
|
| |
|
Event or Problem Description
|
|
Primary procedure, left tha.It was reported that when the size 60 shell was opened, the shell had come off the post that holds it in place in the packaging and some of the outer beading was abraded off the shell.A size 62 shell was opened, and was also loose in the packaging with some of the outer beading removed.The surgeon decided to go with a revision shell which was available in the operating room surgery was completed successfully with a delay of less than a minute.
|
| |
|
Additional Manufacturer Narrative
|
|
An event regarding a packaging issue involving a restoration shell was reported.The event was confirmed following visual inspection of the returned part.Method & results: device evaluation and results: visual inspection was performed and it was identified that there was damage caused to the blister walls and the implant.Material analysis, functional and dimensional inspections were not completed as these aspects are not in question.Clinician review: no medical records were received for review with a clinical consultant.Device history review: all devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the reported lot.Conclusion: the packaging issue was confirmed following visual inspection of the returned part where it was identified that there was damage caused to the blister walls and the implant.Nc was initiated to evaluate the root cause of the event.Please see nc for further information.
|
| |
|
Event or Problem Description
|
|
Primary procedure, left tha.It was reported that when the size 60 shell was opened, the shell had come off the post that holds it in place in the packaging, and some of the outer beading was abraded off the shell.A size 62 shell was opened and was also loose in the packaging with some of the outer beading removed.The surgeon decided to go with a revision shell which was available in the o.R.Surgery was completed successfully with a delay of less than a minute.
|
| |
|