| Catalog Number |
0940023000 |
| Medical Device Problem Code |
Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
Abrasion (1689)
|
| Date of Event |
05/14/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A follow up report will be filed once the quality investigation is complete.
|
| |
|
Event or Problem Description
|
|
It was reported that following the removal of a cast it was observed that the customer was cut or burned.No further information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation results indicate that excessive use of the blade along with using the handpiece in a continuous linear motion may have contributed to the abrasion.The ifu for cast cutter blades has the following warning; ¿cut with an up and down motion.Straight cutting may cut or burn the patient.¿ the quality investigation is complete.
|
| |
|
Event or Problem Description
|
|
It was reported that following the removal of a cast it was observed that the customer was cut or burned.No further information was provided.
|
| |
|