• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SOFRADIM PRODUCTION SAS MESH SOFRADIM - PARIETEX¿ COMPOSITE MESH; MESH, SURGICAL, POLYMERIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SOFRADIM PRODUCTION SAS MESH SOFRADIM - PARIETEX¿ COMPOSITE MESH; MESH, SURGICAL, POLYMERIC Back to Search Results
Model Number PCO2H3
Medical Device Problem Codes Material Erosion (1214); Migration or Expulsion of Device (1395); Product Quality Problem (1506); Adverse Event Without Identified Device or Use Problem (2993); Patient Device Interaction Problem (4001)
Health Effect - Clinical Codes Adhesion(s) (1695); Erosion (1750); Unspecified Infection (1930); Inflammation (1932); Nausea (1970); Vomiting (2144); Hernia (2240); Injury (2348); No Code Available (3191); Unspecified Tissue Injury (4559); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
The patient¿s attorney alleged a deficiency against the device.The product was used for therapeutic treatment of a hernia.It was reported that after implant, the patient experienced an adverse outcome.
 
Event or Problem Description
The patient¿s attorney alleged a deficiency against the device.The product was used for therapeutic treatment of a hiatal hernia in a laparoscopic repair.It was reported that after implant, the patient experienced recurrence, adhesions, gastric fundus diverticulum, nausea, vomiting, adhesions, inflammation, chronically eroded black mesh and mesh erosion.Post-operative patient treatment included lysis of adhesions, small bowel resection and recurrent hiatal hernia repair.
 
Additional Manufacturer Narrative
This information was received as a part of an extensive mesh litigation submission to medtronic.The fda was notified of this large complaint receipt.Due to the volume of complaint information received by medtronic, this resulted in a report beyond the 30 day target.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
This information was received as a part of an extensive mesh litigation submission to medtronic.The fda was notified of this large complaint receipt.Due to the volume of complaint information received by medtronic, this resulted in a report beyond the 30 day target.Additional information: a4(weight in lbs), b2(change other to intervention required), b5, b7, d7, g4, h6 medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
The patient¿s attorney alleged a deficiency against the device.The product was used for therapeutic treatment of a hiatal hernia.It was reported that after implant, the patient experienced recurrence, adhesions and mesh erosion.Post-operative patient treatment included lysis of adhesions, small bowel resection and recurrent hiatal hernia repair.
 
Event or Problem Description
The patient¿s attorney alleged a deficiency against the device.The product was used for therapeutic treatment of a hiatal hernia in a laparoscopic repair.It was reported that after implant, the patient experienced recurrence, adhesions, gastric fundus diverticulum, nausea, vomiting, adhesions, inflammation, chronically eroded black mesh, infection, mesh migration and mesh erosion.Post-operative patient treatment included lysis of adhesions, small bowel resection and recurrent hiatal hernia repair.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
The patient¿s attorney alleged a deficiency against the device.The product was used for therapeutic treatment of a hiatal hernia in a laparoscopic repair.It was reported that after implant, the patient experienced recurrence, adhesions, gastric fundus diverticulum, nausea, vomiting, adhesions, inflammation, chronically eroded black mesh, infection, mesh migration and mesh erosion.Post-operative patient treatment included medication, removal of mesh, ct scan, lysis of adhesions, small bowel resection and recurrent hiatal hernia repair.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MESH SOFRADIM - PARIETEX¿ COMPOSITE MESH
Common Device Name
MESH, SURGICAL, POLYMERIC
Manufacturer (Section D)
SOFRADIM PRODUCTION SAS
116 avenue du formans
trevoux 01600
FR  01600
Manufacturer (Section G)
SOFRADIM PRODUCTION SAS
116 avenue du formans
trevoux 01600
FR   01600
Manufacturer Contact
justin ellis
5920 longbow drive
8200 coral sea st ne
mounds view,mn, CT 55112
7635265677
MDR Report Key8687468
Report Number9615742-2019-02019
Device Sequence Number19426299
Product Code FTL
UDI-Device Identifier10884521178465
UDI-Public10884521178465
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K040998
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Device Explanted Year2016
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date (Section B) 05/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2017
Device Model NumberPCO2H3
Device Catalogue NumberPCO2H3
Device Lot NumberPME00227
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 05/15/2019
Supplement Date Received by Manufacturer05/15/2019
05/15/2019
12/22/2021
05/10/2023
Initial Report FDA Received Date06/11/2019
Supplement Report FDA Received Date11/08/2019
12/16/2019
01/09/2022
05/12/2023
Date Device Manufactured05/27/2012
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNK PROTACK(LOT#: UNKNOWN)
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient SexFemale
Patient Weight60 KG
-
-