MAUDE Adverse Event Report: DEPUY IRELAND 9616671 ATTUNE RP TIB BASE SZ 2 CEM; ATTUNE IMPLANT : KNEE TIBIAL TRAY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND 9616671 ATTUNE RP TIB BASE SZ 2 CEM; ATTUNE IMPLANT : KNEE TIBIAL TRAY Back to Search Results
Catalog Number 150610003
Medical Device Problem Code Loss of or Failure to Bond (1068)
Health Effect - Clinical Codes Pain (1994); No Code Available (3191)
Date of Event 05/20/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.(b)(4).
 
Event or Problem Description
The patient was operated on about a year ago and recently underwent a repeat operation due to suspicion of infection.When the knee opened towards the first stage, the knee did not appear to be contaminated at all, but the tibial surface was completely loose and was removed with the finger when there was no connection between the cement and the implant.As for the connection between the cement and the bone it seems that there is an excellent connection.As a result, it is likely that the cause of the patient's pain is not the infection but the weakening of the tibial surface.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary
=
> no device was received.Root cause undetermined.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot
=
> production code 150610003, work order 8619504 was manufactured on 13 september 2017.(b)(4) manufactured per specification and all raw materials met specification.There were no ncs or deviations associated with this lot.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE RP TIB BASE SZ 2 CEM
Common Device Name
ATTUNE IMPLANT : KNEE TIBIAL TRAY
Manufacturer (Section D)
DEPUY IRELAND 9616671
loughbeg ringaskiddy co.
cork IN
EI 
MDR Report Key8687999
Report Number1818910-2019-95727
Device Sequence Number2349161
Product Code NJL
UDI-Device Identifier10603295042563
UDI-Public10603295042563
Combination Product (Y/N)N
Initial Reporter CountryIS
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2019
Reporter Type Manufacturer
Report Source foreign,health professional,l
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/21/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number150610003
Device Lot Number8619504
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 05/21/2019
Supplement Date Received by Manufacturer07/14/2019
08/05/2019
Initial Report FDA Received Date06/11/2019
Supplement Report FDA Received Date07/22/2019
08/06/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-