| Catalog Number |
25002112E |
| Medical Device Problem Code |
Device Damaged by Another Device (2915)
|
| Health Effect - Clinical Code |
No Information (3190)
|
| Date of Event |
05/13/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report is for one of two devices involved in this event, please refer to report 3013450937-2019-00023.The device history record and sterilization batch release record were reviewed and found that the design manufacturing and sterilization did not lead to this complaint.The component was disposed of and was unable to be further investigated.Should additional information be obtained the report will be supplemented.
|
| |
|
Event or Problem Description
|
|
An eleos resurfacing femur limb salvage system was implanted and while assembling the components, the threads of the axial pin inserter/extractor fractured off within the resurfacing femur axial pin.The resurfacing femur axial pin was removed, and a new pin was placed.
|
| |
|
Event or Problem Description
|
|
An eleos resurfacing femur limb salvage system was implanted and while assembling the components, the threads of the axial pin inserter/extractor fractured off within the resurfacing femur axial pin.The resurfacing femur axial pin was removed, and a new pin was placed.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is for one of two devices involved in this event, please refer to report 3013450937-2019-00023.The device history record and sterilization batch release record were reviewed and found that the design manufacturing and sterilization did not lead to this complaint.The component was disposed of and was unable to be further investigated.Should additional information be obtained the report will be supplemented.
|
| |
|