(b)(4).Device evaluation summary: an unopened sample of product code vcp215, lot pbz938 was returned for analysis.During the visual inspection of the sample, no defects were found on the package.The sample was opened, and the swage and attachment area were noted to be as expected.The suture was dispensed without problem and examined along of the strand and no defects, fraying suture or diameter issue were noted.A functional test was performed to strands using a federal gauge to measure the suture diameter and the result meet the requirements.Per the condition of the representative sample, no performance fraying suture intra-op or suture diameter was found and the tested sample met the finished goods requirements.A manufacturing record evaluation was performed for the finished device lot number pbz938 and no non-conformances were identified.
|