HOYA SURGICAL OPTICS, INC. HOYA ISERT 251; INTRAOCULAR LENS Back to Search Results
Model Number HOYA ISERT 251 (+23.00 D)
Medical Device Problem Code Scratched Material (3020)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 06/24/2015
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
This report is being submitted after the initial 30-day reporting period deadline; it is part of a corrective action being implemented per internal document capa-(b)(4) for outside us like products reporting.This initial and final emdr is being submitted to fda for outside us like products reporting.The device was returned to the manufacturer, and the product investigation was conducted.The results are listed below: the lens was ejected out from the injector.Upon closer inspection at 20x magnification, there was sign of scratches observed on the surface of the lens.No abnormalities were found in production and inspection records of the product.Exact root cause is not determined.Risk analysis conducted on the product line and failure code is noted below: a review of the most recent complaint trending data indicates that no significant trends have been identified at this time, and no capa is required as part of the product evaluation.
 
Event or Problem Description
Injector was filled as usual with methocel.The iol could be easily advanced.Implantation as usual - after the lens did fold punctual deepening (wrinkle) on the lens was observed.Iol was explanted.Another lens has been implanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOYA ISERT 251
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
HOYA SURGICAL OPTICS, INC.
15335 fairfield ranch rd.
suite 250
chino hills CA 91709
MDR Report Key8689652
Report Number3006723646-2019-00060
Device Sequence Number2511149
Product Code HQL
UDI-Device Identifier04547480450353
UDI-Public04547480450353
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
P080004
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2015
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 07/24/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2018
Device Model NumberHOYA ISERT 251 (+23.00 D)
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/03/2015
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/24/2015
Initial Report FDA Received Date06/11/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-