• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HOYA SURGICAL OPTICS, INC. HOYA ISERT 250; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOYA SURGICAL OPTICS, INC. HOYA ISERT 250; INTRAOCULAR LENS Back to Search Results
Model Number HOYA ISERT 250 (+16.50 D)
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 02/29/2016
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
This report is being submitted after the initial 30-day reporting period deadline; it is part of a corrective action being implemented per internal document capa-(b)(4) for outside us like products reporting.This initial and final emdr is being submitted to fda for outside us like products reporting.The device was not returned to the manufacturer; therefore, the product investigation consisted of a review of the manufacturing records.The results are listed below: no abnormalities were found in the production and inspection records of the product.Exact root cause is not determined.Risk analysis conducted on the product line and failure code is noted below: a review of the most recent complaint trending data indicates that no significant trends have been identified at this time, and no capa is required as part of the product evaluation.
 
Event or Problem Description
Torn haptic - must be explanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOYA ISERT 250
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
HOYA SURGICAL OPTICS, INC.
15335 fairfield ranch rd.
suite 250
chino hills CA 91709
MDR Report Key8689706
Report Number3006723646-2019-00076
Device Sequence Number3029710
Product Code HQL
UDI-Device Identifier04547480480718
UDI-Public04547480480718
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
P080004
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2016
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 02/29/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2018
Device Model NumberHOYA ISERT 250 (+16.50 D)
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/29/2016
Initial Report FDA Received Date06/11/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-