| Model Number |
86700 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Cardiopulmonary Arrest (1765); Low Blood Pressure/ Hypotension (1914)
|
| Date of Event |
05/28/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient experienced hypotension.During an ablation procedure for atrial flutter, a blazer dx-20 catheter was inserted in the patient with no issues.Then, an intellamap orion catheter was inserted into the patient with no issues.About 5 minutes later, the patient's blood pressure dropped.They had to do compressions and then the patient's pressure was fine.The case was then cancelled.The doctor checked with ultrasound to check if there was an effusion from the catheters and there was not.The physician believed that the event was due an anesthesia issue and not the devices.The patient was fine and there were no other complications.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient experienced hypotension.During an ablation procedure for atrial flutter, a blazer dx-20 catheter was inserted in the patient with no issues.Then, an intellamap orion catheter was inserted into the patient with no issues.About 5 minutes later, the patient's blood pressure dropped.They had to do compressions and then the patient's pressure was fine.The case was then cancelled.The doctor checked with ultrasound to check if there was an effusion from the catheters and there was not.The physician believed that the event was due an anesthesia issue and not the devices.The patient was fine and there were no other complications.It was further reported that the "compressions" meant the patient required cardiopulmonary resuscitation (cpr).
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned for analysis.The device did not have visual defects.The steering knob and the tension control knob functioned properly on both lock and unlock positions.No abnormal resistance was felt when actuating the steering mechanism.The right and left curves are placed in the template shaded areas, the device passed the dimensional test.The electrical test was performed and the device was found within specification.No opens or shorts were found.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.
|
| |
|
Search Alerts/Recalls
|