MAUDE Adverse Event Report: MEDTRONIC NEUROSURGERY CSF-FLOW CONTROL VALVE, BURR HOLE, 16 MM, LOW PRESSURE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NEUROSURGERY CSF-FLOW CONTROL VALVE, BURR HOLE, 16 MM, LOW PRESSURE Back to Search Results
Model Number 42542
Medical Device Problem Codes Break (1069); Fracture (1260); Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 06/10/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Information was received regarding a patient undergoing csf management surgery.It was reported that when pre-testing to vent the product, the device was unable to absorb water, the whole device flattened, and it could not be exhausted.There were no related patient symptoms.
 
Additional Manufacturer Narrative
The returned valve met the requirements for patency, leak, reflux, pressure-flow and preimplantation testing.Therefore, the nature of the compliant could not be replicated by the laboratory personnel.Crystalline debris was observed on the interior of the valve.The instructions for use cautions, ¿care must be taken to ensure that particulate contaminates are not introduced into the shunt components during preimplantation testing or handling.Introduction of contaminates could result in improper performance (overdrainage or underdrainage) of the shunt system.Particulate matter, which enters the shunt system, may also hold pressure/flow controlling mechanisms open resulting in overdrainage.¿ all valves are 100% tested at the time of manufacture.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CSF-FLOW CONTROL VALVE, BURR HOLE, 16 MM, LOW PRESSURE
Manufacturer (Section D)
MEDTRONIC NEUROSURGERY
125 cremona drive
goleta CA 93117
MDR Report Key8691855
Report Number2021898-2019-00229
Device Sequence Number2073386
Product Code JXG
UDI-Device Identifier00643169466043
UDI-Public00643169466043
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K831678
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2021
Device Model Number42542
Device Catalogue Number42542
Device Lot NumberE10606
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/26/2019
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/11/2019
Supplement Date Received by Manufacturer06/28/2019
Initial Report FDA Received Date06/12/2019
Supplement Report FDA Received Date07/19/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-