• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 MAGIC3 GO MALE CATHETER; MAGIC 3 GO MALE COUDE CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 MAGIC3 GO MALE CATHETER; MAGIC 3 GO MALE COUDE CATHETER Back to Search Results
Model Number 50818G
Medical Device Problem Code Device Damaged Prior to Use (2284)
Health Effect - Clinical Codes Abrasion (1689); Injury (2348)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event or Problem Description
It was reported that the coude tip had spurs at the end which allegedly cut the patient during use of the catheter.No medical intervention was required.
 
Additional Manufacturer Narrative
The reported event was confirmed as manufacturing related.Visual evaluation noted one magic 3 go intermittent catheter received.Visual evaluation noted one large sharp spur was on the tip of the catheter.A potential root cause for this failure could be mandrels touching each other within the pallet.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: ¿review the self-catheterization procedure with a healthcare professional.1.Always wash hands prior to use.2.Open the catheter package by peeling the tab upwards with the aid of the finger hole.3.If desired, use the adhesive label(s) to hang the package on a nearby dry vertical surface while preparing to catheterize.4.Clean the opening to the urethra ¿ and the surrounding area.Wipe from front to back to avoid fecal contamination.Wash your hands again.5.Hold the insertion sleeve with your dominant hand and squeeze it to grip the catheter shaft as you remove the catheter from the pack.Next, hold the catheter funnel above the insertion sleeve with your other hand and slide the insertion sleeve down the shaft, stopping at about 6¿ from the tip.Release the funnel.Using the insertion sleeve to hold the catheter firmly, gently pass the tip of the catheter into your urethra until the insertion sleeve nears the meatus.Repeat until urine starts to flow.6.Keep the catheter steady until urine stops flowing.When urine stops flowing, slowly withdraw the catheter, stopping if flow starts again, until the last few drops have drained.Check the color, odor and clarity of the urine.Any changes may need to be reported to a health care professional.Disposal 7.Place the catheter back into the package and dispose in accordance with accepted medical practice and applicable local, state and federal laws and regulations.Do not flush down the toilet.8.Wash hands.".
 
Event or Problem Description
It was reported that the coude tip had spurs at the end which allegedly cut the patient during use of the catheter.No medical intervention was required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGIC3 GO MALE CATHETER
Common Device Name
MAGIC 3 GO MALE COUDE CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key8691911
Report Number1018233-2019-03077
Device Sequence Number11334341
Product Code EZD
UDI-Device Identifier00801741137044
UDI-Public(01)00801741137044
Combination Product (Y/N)N
Initial Reporter StateLA
Initial Reporter CountryUS
PMA/510(K) Number
K172247
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other,user facility
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 07/23/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date09/28/2021
Device Model Number50818G
Device Catalogue Number50818G
Device Lot NumberJUCW0235
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/14/2019
Initial Date Received by Manufacturer 05/23/2019
Supplement Date Received by Manufacturer07/10/2019
Initial Report FDA Received Date06/12/2019
Supplement Report FDA Received Date07/23/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age72 YR
-
-