Viabahn stent-graft being used to treat stenosis of av graft outflow vein.The stent-graft was situated perfectly in a straight vane segment, over a 0.038" bentson guidewire, which is our standard for this procedure.During attempted deployment, after approximately 2/3 of the stent have been expanded, deployment dashboard is suddenly struck and could not be pulled any further.Multiple attempts to get the court to release the remaining unopened stent-graft were unsuccessful.The wire was replaced with a super-stiff amplatz wire, which did not substantially change the configuration of the stent-graft; the cord would still not release the stent-graft.There was marked deformation and segmental particle expansion of the stent-graph.Finally, with intentionally excessive force, the [invalid] snapped, and the stent was left completely opened, resulting in obstruction of the target vessel.Attempts to extract the stent using a snare were unsuccessful.The graft had to be occluded to avoid extravasation around the mal-deployed sent.The patient was sent by ambulance to hospital, admitted for a management of this complication, observation, and surgical graft salvage.He needed acute femoral hd catheter for treatment of renal conditions (hyperkalemia, hypertension, fluid overload).Surgery was not willing to operate on a graft due to the potential difficulty reaching healthy target made.A long-term tunneled hemodialysis catheter was placed, and he was discharged with plans for a new av access creation.The otherwise good quality dialysis graft had to be sacrificed.Note that we have placed several hundreds of the viabahn stent-grafts in similar situations.On rare occasion, we had the deployed cord "stick" and meet some resistance; each time we have been able to overcome back with additional gentle traction.This is the first time we have had a deployment failure.Fda safety report id # (b)(4).
|