• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC NEUROMODULATION; PROBE, RADIOFREQUENCY LESION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC NEUROMODULATION; PROBE, RADIOFREQUENCY LESION Back to Search Results
Model Number CSK-TC15
Medical Device Problem Code Defective Device (2588)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 05/23/2019
Type of Reportable Event Malfunction
Event or Problem Description
Device analysis performed on the returned electrode found that the shaft of the electrode was separated from the hub.The damage was likely due to unintended excessive external mechanical force.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NA
Common Device Name
PROBE, RADIOFREQUENCY LESION
Manufacturer (Section D)
BOSTON SCIENTIFIC NEUROMODULATION
25155 rye canyon loop
valencia CA 91355
MDR Report Key8692499
Report Number3006630150-2019-02753
Device Sequence Number13866737
Product Code GXI
UDI-Device Identifier00813250010114
UDI-Public00813250010114
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K050084
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 06/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberCSK-TC15
Device Catalogue NumberCSK-TC15
Device Lot NumberM063
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/12/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/23/2019
Initial Report FDA Received Date06/12/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-