| Model Number |
N/A |
| Medical Device Problem Code |
Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zimmer biomet complaint (b)(4).Concomitant medical products: biomet microfixation ratcheting screwdriver catalog #: 46-0008; lot #: ni.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the driver is stripped.Attempts have been made and no further information has been provided.No adverse events have been reported as a result of the malfunction.
|
| |
|
Event or Problem Description
|
|
This follow-up report is being submitted to relay additional information.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The complaint is confirmed.On visual inspection, the contra angle driver appears to be in fair overall shape with minor discoloration noted.When rotating the bottom of the driver, there wasn't any significant sticking or resistance noted.An sp-2379 lot# 683420 contra angle driver blade was inserted into the collet and the assembly was functionally tested by attempting to drive a 2.4mm cancellous locking screw into a block of white oak wood.When the screw was driven approximately halfway, the driver was noted to stop functioning and the screw would not advance any further when rotating the gear.The driver was disassembled for further inspection and when disassembling the collet there were metal shavings noted to fall out from the assembly and the teeth of the gear were worn.The dhr was reviewed and there were no non-conformances found.There are no indications of manufacturing defects.The most likely underlying cause of the complaint is the gears experienced force in excess of what they were designed to encounter when driving the screw.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Search Alerts/Recalls
|