It was reported that during a sialendoscopy procedure to remove salivary gland stones, the salivary stone extractor basket broke under pressure.The basket was "being used heavily for removing the stone" from the patient before it broke.Another salivary stone extractor basket was opened to complete the procedure.As reported, no adverse effects were experienced by the patient due to this occurrence.
|
|
Blank fields on this form indicate the information is unknown, unchanged or unavailable.Investigation - evaluation.A review of documentation including the complaint history, device history record, instructions for use, manufacturing instructions and quality control of the device was conducted during the investigation.The complaint device was not returned; therefore, a physical examination of the device not conducted.However, a document-based investigation evaluation was performed.It was concluded that the device aspect in question was visually and functionally inspected by quality control and no related gaps in production or processing controls were noted.The risk specification for this product includes the loss of use failure mode and identifies the risk controls that are in place to mitigate the risk of this type of failure.A review of the device history record showed no nonconformances in lot 9548126 that could have contributed to the failure of the device.It should be noted that there have been no other complaints reported for this lot number.There is no evidence to suggest that nonconforming exists in house or in the field.The instructions for use (ifu), provides the following information to the user related to the reported failure mode: precaution.Enclose the device in the sheath before removing from the tray/holder.Do not use excessive force to manipulate this device.Damage to the device may occur.Based on the information provided, no product returned and the results of our investigation, a definitive root cause could not be established.It was reported that the complaint device, ¿was used heavily for removing the stone and broke under pressure.¿ it appears likely that the device broke from stresses placed on it during use.Per the quality engineering risk assessment no further action is required.The appropriate personnel have been notified and we will continue to monitor for similar complaints.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|