A steris service technician arrived onsite following the reported event to inspect the amsco 400 sterilizer.The technician was informed that the employee subject of the reported event had hit the bottom panel with a cart, causing the panel to fall off and the reported event to occur.The panel located on the bottom of the amsco 400 sterilizer is held on by magnets.Forceful impact to the panel can cause it to detach from the sterilizer and fall.The technician re-secured the panel back onto the sterilizer, tested the unit, and confirmed it to be operating according to specification.The unit was returned to service.No additional issues have been reported.
|