MEDTRONIC HEART VALVES DIVISION CG FUTURE BAND; RING, ANNULOPLASTY Back to Search Results
Model Number 638RL28
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Corneal Pannus (1447); Mitral Valve Stenosis (1965); No Information (3190)
Date of Event 05/30/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product analysis: no product was returned.Conclusion: without the return of the product, no definitive conclusion can be made regarding the clinical observation.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received information that 8 years and 4 months post implant of this mitral annuloplasty ring, the ring was explanted and replaced with a bioprosthetic valve. the reason for the explant was not reported.  no additional adverse patient effects were reported. .
 
Event or Problem Description
Medtronic received information that 8 years and 4 months post implant of this mitral annuloplasty ring, the ring was explanted and replaced with a bioprosthetic valve. the reason for the explant was not reported.  no additional adverse patient effects were reported.  additional information was received that the reason for replacement was stenosis.It was also noted that there was "severe significant pannus ingrowth" on the patient's native mitral valve leaflets.No additional adverse patient effects were reported.
 
Additional Manufacturer Narrative
Medtronic received additional information that the reason for replacement was stenosis.It was also noted that there was "severe si gnificant pannus ingrowth" on the patient's native mitral valve leaflets.No additional adverse patient effects were reported.Updated patient code to field h6.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CG FUTURE BAND
Common Device Name
RING, ANNULOPLASTY
Manufacturer (Section D)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
MDR Report Key8693330
Report Number2025587-2019-01908
Device Sequence Number1941471
Product Code KRH
UDI-Device Identifier00613994751720
UDI-Public00613994751720
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K061127
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2011
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/06/2015
Device Model Number638RL28
Device Catalogue Number638RL28
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/06/2019
Supplement Date Received by Manufacturer07/05/2019
Initial Report FDA Received Date06/12/2019
Supplement Report FDA Received Date07/25/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age74 YR
-
-