| Model Number |
CD1275-36Q |
| Medical Device Problem Code |
Defibrillation/Stimulation Problem (1573)
|
| Health Effect - Clinical Codes |
Hypoxia (1918); Neurological Deficit/Dysfunction (1982)
|
| Date of Event |
05/27/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined. .
|
| |
|
Event or Problem Description
|
|
The patient experienced ventricular fibrillation (vf).Therapy was delivered but was not successful in terminating the vf.The patient received external defibrillation which finally terminated the arrhythmia.Due to the length of the episode, the patient suffered cerebral consequences due to hypoxia.The patient was sent to the intensive care unit (icu) and is on a respirator.There is no allegation of malfunction against the device.Additional information has been requested and not yet received.
|
| |
|
Event or Problem Description
|
|
Additional information was received indicating the device was functioning as programmed and there was no alleged malfunction of the device.Furthermore, the device was unable to terminate the ventricular fibrillation due to patient condition.
|
| |
|