• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MASIMO - 40 PARKER M-LNCS TC-I; OXIMETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MASIMO - 40 PARKER M-LNCS TC-I; OXIMETER Back to Search Results
Model Number 2503
Medical Device Problem Code Incorrect Measurement (1383)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/06/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product has been returned to the local facility but has not yet been received at the main office for evaluation.Once returned and investigated, a follow-up report will be submitted.Initial reporter telephone number exceeded the maximum allowable characters, the telephone number is as follows: (b)(6).
 
Event or Problem Description
The customer reported inaccurate values.No consequences or impact to patient were reported.
 
Additional Manufacturer Narrative
Additional manufacturing narrative: other, other text: the returned sensor was evaluated.During evaluation, the cable passed visual inspection; eeprom contents were verified and no problems were observed through this verification.  during continuity testing, the cable failed due to a defective emitter causing low voltage on the red diode.The sensor is functioning, however the low voltage on the red diode caused the spo2 values to fluctuate from 43 to 99%.No stable spo2 measurement value was provided.No issues were detected with regard to pulse rate values.Initial reporter telephone number exceeded the maximum allowable characters, the telephone number is as follows: (b)(6).
 
Event or Problem Description
The customer reported inaccurate values.No consequences or impact to patient were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
M-LNCS TC-I
Common Device Name
OXIMETER
Manufacturer (Section D)
MASIMO - 40 PARKER
40 parker
irvine CA 92618 1604
MDR Report Key8694190
Report Number2031172-2019-00283
Device Sequence Number1748157
Product Code DPZ
UDI-Device Identifier00843997003230
UDI-Public00843997003230
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K051212
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 05/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number2503
Device Catalogue Number2503
Device Lot Number18FQZ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/27/2019
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 05/20/2019
Supplement Date Received by Manufacturer06/24/2019
Initial Report FDA Received Date06/12/2019
Supplement Report FDA Received Date07/15/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-