| Model Number |
2503 |
| Medical Device Problem Code |
Incorrect Measurement (1383)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/06/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The product has been returned to the local facility but has not yet been received at the main office for evaluation.Once returned and investigated, a follow-up report will be submitted.Initial reporter telephone number exceeded the maximum allowable characters, the telephone number is as follows: (b)(6).
|
| |
|
Event or Problem Description
|
|
The customer reported inaccurate values.No consequences or impact to patient were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional manufacturing narrative: other, other text: the returned sensor was evaluated.During evaluation, the cable passed visual inspection; eeprom contents were verified and no problems were observed through this verification. during continuity testing, the cable failed due to a defective emitter causing low voltage on the red diode.The sensor is functioning, however the low voltage on the red diode caused the spo2 values to fluctuate from 43 to 99%.No stable spo2 measurement value was provided.No issues were detected with regard to pulse rate values.Initial reporter telephone number exceeded the maximum allowable characters, the telephone number is as follows: (b)(6).
|
| |
|
Event or Problem Description
|
|
The customer reported inaccurate values.No consequences or impact to patient were reported.
|
| |
|