• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TYRX, INC. AIGIS-R LARGE; MESH, SURGICAL, POLYMERIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TYRX, INC. AIGIS-R LARGE; MESH, SURGICAL, POLYMERIC Back to Search Results
Model Number CMRM6133
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 04/11/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the implantable pulse generator (ipg) system was removed due to an infection.The patient was having an upgrade procedure from an ipg to an implantable cardioverter defibrillator (icd) and when the physician opened the pocket, yellow drainage was noticed in the pocket.Cultures were taken.No further patient complications have been reported as a result of this event.It was further reported that the implantable pulse generator (ipg) had been implanted with an absorbable envelope.It was also further reported that the aerobic culture tested positive for propionibacterium acnes.
 
Additional Manufacturer Narrative
Correction to g4 on initial report: should have been 30 may 2019.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIGIS-R LARGE
Common Device Name
MESH, SURGICAL, POLYMERIC
Manufacturer (Section D)
TYRX, INC.
1 deer park dr.
suites g, e
monmouth jct NJ 08852
MDR Report Key8694706
Report Number3005619263-2019-00022
Device Sequence Number14071649
Product Code FTL
UDI-Device Identifier00643169468795
UDI-Public00643169468795
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K130943
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date11/30/2016
Device Model NumberCMRM6133
Device Catalogue NumberCMRM6133
Device Lot Number16K21787
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/04/2017
Supplement Date Received by Manufacturer02/29/2020
Initial Report FDA Received Date06/13/2019
Supplement Report FDA Received Date02/29/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
A2DR01 IPG, 5076-52 LEAD, 5076-45 LEAD; A2DR01 IPG, 5076-52 LEAD, 5076-45 LEAD
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age66 YR
Patient Weight102
-
-