| Model Number |
60SF2 |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
05/23/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Return of the device has been requested from the customer and is currently pending confirmation of return.Investigation is ongoing.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information that during use of the 60sf2 cryocath probe, after performing 2 one-minute freezes and 1 two-minute freeze, the probe broke and fell into the patient.The customer had used warm saline to defrost the probe.The probe was retrieved by forceps with no tissue damage.There was no difficulty retrieving the broken piece of the probe.No adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: the bellows have been manipulated beyond the design and or material capabilities.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information that during use of the 60sf2 cryocath probe, after performing 2 one-minute freezes and 1 two-minute freeze, the probe broke and fell into the patient.The customer had used warm saline to defrost the probe.The probe was retrieved by forceps with no tissue damage.There was no difficulty retrieving the broken piece of the probe.No adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Visual analysis of the returned device confirmed a section of the heat-exchanger broken off.The investigation is currently ongoing.Following completion of the investigation, conclusion will be provided.If information is provided in the future, a supplemental report will be issued.
|
| |
|