• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDACTA INTERNATIONAL SA CUP: VERSAFITCUP CC TRIO ACETABULAR SHELL CC TRIO Ø 52; CEMENTLESS ACETABULAR SHELL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDACTA INTERNATIONAL SA CUP: VERSAFITCUP CC TRIO ACETABULAR SHELL CC TRIO Ø 52; CEMENTLESS ACETABULAR SHELL Back to Search Results
Catalog Number 01.26.45.0052
Medical Device Problem Code Malposition of Device (2616)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 05/14/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Batch review performed on 07 jun 2019: lot 164700: (b)(4) items manufactured and released on 16-nov-2016.Expiration date: 2021-11-03.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been already sold without any similar reported event.
 
Event or Problem Description
About 1 year after primary surgery the surgeon revised the medacta total hip due to acetabulum implant malposition, approximatively 50° anteverted, which caused cup and stem to rub on each other and stem had worked loose.All implants revised successfully and competitor implants implanted.Primary surgery performed with amis technique.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CUP: VERSAFITCUP CC TRIO ACETABULAR SHELL CC TRIO Ø 52
Common Device Name
CEMENTLESS ACETABULAR SHELL
Manufacturer (Section D)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, 6874
SZ  6874
MDR Report Key8695353
Report Number3005180920-2019-00458
Device Sequence Number16809930
Product Code LZO
UDI-Device Identifier07630030807770
UDI-Public07630030807770
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K103352
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 06/13/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/03/2021
Device Catalogue Number01.26.45.0052
Device Lot Number164700
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/14/2019
Initial Report FDA Received Date06/13/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age54 YR
Patient Weight75
-
-