| Model Number |
2477-0000 |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Tear, Rip or Hole in Device Packaging (2385)
|
| Health Effect - Clinical Code |
No Information (3190)
|
| Date of Event |
02/01/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A 100ml baxter bag ndc 0338-0049-48 lot p375030 exp aug19 0.9% sodium chloride injection and one used hemostat.Pediatric patient.(picu).An unexpected return was received from the customer without details or information of the event.Visual inspection confirmed a leak at the bottom of the silicone pump segment (p/n 12088541).The set was received with blood throughout the tubing.Functional testing identified a leak at the same location.Closer inspection under a lab microscope observed that the leak was due to a small tear on the silicone segment.No tool marks or crush marks were observed on the lower fitment near the tear.The cause of the leak was a tear in the silicone section of the tubing.The root cause of the tear was not identified.
|
| |
|
Event or Problem Description
|
|
The customer reported that supply chain received a set from the picu; the failure mode with the set was not provided.Although requested, no additional information was available.
|
| |
|