• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BD CARIBE LTD. BD MICROTAINER® SST¿ - AMBER; BLOOD SPECIMEN COLLECTION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD CARIBE LTD. BD MICROTAINER® SST¿ - AMBER; BLOOD SPECIMEN COLLECTION DEVICE Back to Search Results
Catalog Number 365978
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 05/29/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Date of event: unknown.The date received by manufacturer has been used for this field.The reported lot # [8236759] was not found for the reported catalog # [365978].Medical device expiration date: unknown.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.Device manufacture date: unknown.
 
Event or Problem Description
It was reported that the bd microtainer® sst¿ - amber wouldn't clot after use.This complaint was created to capture the 2nd of 2 related incidents.The following information was provided by the initial reporter: "it was reported the customer was experiencing tubes not clotting.Found out about the recall after experiencing issues.This has been occurring for a couple weeks, no specific dates known.".
 
Additional Manufacturer Narrative
H.6.Investigation summary: bd had not received samples or photos from the customer facility for evaluation; therefore, the investigation was limited.Additionally, bd was unable to determine the specific lot number associated with this complaint.Therefore, a review of the device history record could not be conducted.As no samples or photos were received for evaluation, the customer's indicated failure mode was not observed by bd.As there was no sample or photo available for evaluation, a root cause could not be determined.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.The bd business team regularly reviews the collected data for identification of emerging trends.
 
Event or Problem Description
It was reported that the bd microtainer® sst¿ - amber wouldn't clot after use.This complaint was created to capture the 2nd of 2 related incidents.The following information was provided by the initial reporter: "it was reported the customer was experiencing tubes not clotting.Found out about the recall after experiencing issues.This has been occurring for a couple weeks, no specific dates known.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD MICROTAINER® SST¿ - AMBER
Common Device Name
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BD CARIBE LTD.
road 31
k.m. 24.3
juncos
MDR Report Key8695757
Report Number2618282-2019-00196
Device Sequence Number2386342
Product Code JKA
UDI-Device Identifier50382903659789
UDI-Public50382903659789
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K991702
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other,user facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number365978
Device Lot NumberUNKNOWN
Initial Date Received by Manufacturer 05/29/2019
Supplement Date Received by Manufacturer05/29/2019
Initial Report FDA Received Date06/13/2019
Supplement Report FDA Received Date07/18/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-