Model Number 7015
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/08/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(6).Returned product consisted of a quantum maverick balloon catheter.The shaft, hypotube, tip and balloon were microscopically and visually examined.The balloon was tightly folded.There were numerous hypotube kinks throughout the device.There was a complete hypotube separation 48cm distal of the strain relief.The hypotube fracture surfaces were ovaled, which suggests the device was kinked prior to separation.Inspection of the remainder of the device presented no other damage or irregularities.
 
Event or Problem Description
Reportable based on device analysis completed on 20-may-2019.It was reported that shaft kink occurred.The target lesion was located in the right coronary artery.A 2.5mmx 15mm quantum maverick balloon catheter was used but shaft kink was noted.The procedure was completed with another of the same device.No patient complications were reported and patient status was stable.However, returned device analysis revealed a hypotube separation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUANTUM MAVERICK
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key8695874
Report Number2134265-2019-06879
Device Sequence Number11216539
Product Code LOX
UDI-Device Identifier08714729392767
UDI-Public08714729392767
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P860019
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/13/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/14/2022
Device Model Number7015
Device Catalogue Number7015
Device Lot Number0023188868
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/01/2019
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/20/2019
Initial Report FDA Received Date06/13/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age72 YR
Patient Weight62
-
-