• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. PENUMBRA SMART COIL; HCG, KRD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. PENUMBRA SMART COIL; HCG, KRD Back to Search Results
Catalog Number 400SMTXSFT0408
Medical Device Problem Code Difficult or Delayed Separation (4044)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/14/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product was implanted and not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report number: 3005168196-2019-01144.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure in the right vertebral artery using penumbra smart coils (smart coils) and a penumbra smart coil detachment handle (handle).During the procedure, the physician advanced a smart coil into the target vessel using a non-penumbra microcatheter and used the handle to detach it; however, the smart coil failed to detach.After several failed detachment attempts using the handle, the physician attempted to manually detach the smart coil by separating the end of the pusher assembly; however, it would not to detach.The physician then made a second attempt to manually detach the smart coil and was able to successfully detach it in the target vessel.The procedure was then completed using additional smart coils, the same handle and microcatheter.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SMART COIL
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key8696133
Report Number3005168196-2019-01143
Device Sequence Number14014989
Product Code HCG
UDI-Device Identifier00814548015392
UDI-Public00814548015392
Combination Product (Y/N)Y
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K160832
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/14/2019
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/11/2023
Device Catalogue Number400SMTXSFT0408
Device Lot NumberF87002
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/14/2019
Initial Report FDA Received Date06/13/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age40 YR
Patient Weight77
Date Report Sent to FDA01/01/2005
-
-