• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. INTELLIS; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. INTELLIS; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF Back to Search Results
Model Number 97715
Device Problem Break (1069)
Patient Problem Sepsis (2067)
Event Date 06/12/2019
Event Type  Injury  
Manufacturer Narrative
Concomitant medical products: product id: 977a260, serial #: (b)(4), implanted: (b)(6) 2018, product type: lead.Product id: 977a260, serial #: (b)(4), implanted: (b)(6) 2018, product type: lead.Other relevant device(s) are: product id: 977a260, serial/lot #: (b)(4), ubd: 02-feb-2022, udi #: (b)(4).Product id: 977a260, serial/lot #: (b)(4), ubd: 02-feb-2022, udi #: (b)(4).If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a healthcare professional (hcp), via a manufacturer representative (rep), regarding a patient with an implantable neurostimulator (ins) for non-malignant pain.It was reported that the patient was hospitalized after becoming ill with sepsis.The rep said a ct and x-ray showed one lead to be fractured.It was unknown what factors may have led to the issue.The rep noted that the patient was out of state in a hospital, it was unknown if surgical intervention was planned and the issue had not been resolved at the time of the report.The rep said they would follow-up with the hcp on jun-25.No further complications were reported.
 
Manufacturer Narrative
Product id 977a260, serial# (b)(4), implanted: (b)(6) 2018, product type: lead.Product id 977a260, serial# (b)(4), implanted: (b)(6) 2018, product type: lead.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLIS
Type of Device
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key8697049
MDR Text Key147950534
Report Number3004209178-2019-11563
Device Sequence Number1
Product Code LGW
UDI-Device Identifier00643169781702
UDI-Public00643169781702
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P840001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/13/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/28/2019
Device Model Number97715
Device Catalogue Number97715
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received08/16/2019
Date Device Manufactured01/31/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Life Threatening; Other;
Patient Age64 YR
-
-