• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL AMPLATZER SEPTAL OCCLUDER; TRANSCATHETER SEPTAL OCCLUDER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL AMPLATZER SEPTAL OCCLUDER; TRANSCATHETER SEPTAL OCCLUDER Back to Search Results
Model Number 9-ASD-018
Medical Device Problem Code Inadequacy of Device Shape and/or Size (1583)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 05/13/2019
Type of Reportable Event Serious Injury
Event or Problem Description
On an unknown date, an 18mm amplatzer septal occluder was implanted.Following the procedure, on (b)(6) 2019, the device was reported to have embolized.On (b)(6) 2019, the device was removed from the patient percutaneously by snaring and re-sheathing.The defect was balloon-sized to 18mm using a 30mm pts sizing balloon (lot number: unknown) and the physician attempted closure with a 22mm amplatzer septal occluder (lot number: 6821072) and 26mm amplatzer septal occluder (lot number: 5562243).Both devices were reported to be mis-sized.Of note, the 26mm amplatzer septal occluder was impinging on the mitral valve.The physician elected to abandoned the procedure and surgical closure is planned for a future time.
 
Additional Manufacturer Narrative
An event of embolization was reported.The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications at the time of commercialization.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Additional Manufacturer Narrative
Supplemental report needed for implant date and anatomical difficulties.Correction: supplemental correction report needed for cause of case abandonment traced to patient anatomy.
 
Event or Problem Description
On (b)(6) 2019, an 18mm amplatzer septal occluder was implanted.Following the procedure, the device was reported to have embolized.On (b)(6) 2019, the device was removed from the patient percutaneously by snaring and re-sheathing.The defect was balloon-sized to 18mm using a 30mm pts sizing balloon (lot number: unknown) and the physician attempted closure with a 22mm amplatzer septal occluder (lot number: 6821072) and 26mm amplatzer septal occluder (lot number: 5562243).Both devices were reported to be mis-sized.Of note, the 26mm amplatzer septal occluder was impinging on the mitral valve.Additionally, the patient had multipls asds and challenging anatomical features which caused the mis-sizing.The physician elected to abandoned the procedure and surgical closure is planned for a future time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER SEPTAL OCCLUDER
Common Device Name
TRANSCATHETER SEPTAL OCCLUDER
Manufacturer (Section D)
ST. JUDE MEDICAL
14901 deveau place
minnetonka MN 55345 2126
MDR Report Key8697322
Report Number2182269-2019-00081
Device Sequence Number2472847
Product Code MLV
UDI-Device Identifier05415067019264
UDI-Public05415067019264
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
P000039
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/14/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/31/2021
Device Model Number9-ASD-018
Device Catalogue Number9-ASD-018
Device Lot Number5475335
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 05/16/2019
Supplement Date Received by Manufacturer06/13/2019
07/16/2019
Initial Report FDA Received Date06/13/2019
Supplement Report FDA Received Date07/12/2019
08/14/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-