| Model Number |
AA61181002 |
| Medical Device Problem Code |
Burst Container or Vessel (1074)
|
| Health Effect - Clinical Code |
No Information (3190)
|
| Date of Event |
02/04/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
After receiving this complaint, we searched for other complaint and we didn't find other complaint on this lot.The balloon burst issue is known and monitored on a monthly basis.Possible root cause is a weakness area in the silicone material of the balloon.No other corrective action will be implemented as the specific trend for balloon burst of this product family is considered acceptable as per the threshold.
|
| |
|
Event or Problem Description
|
|
According to the available information, the balloon burst, and catheter came out from patient.Risk of infection and rest of balloon in the bladder.
|
| |
|
Search Alerts/Recalls
|