• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. SILK SUTURE 18"(45CM) 3-0 BLK; SUTURE, NONABSORBABLE, SILK

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. SILK SUTURE 18"(45CM) 3-0 BLK; SUTURE, NONABSORBABLE, SILK Back to Search Results
Catalog Number 632G
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/01/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).A manufacturing record evaluation was performed for the finished device lot number mjh223-632g, and no non-conformances were identified.Investigation summary: unopened samples of product code 632, lot mjh223 was returned for analysis.The issue sample was not received for evaluation.During the visual inspection of samples, no defects were found on the packages.The samples were opened, and the swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along of the strands and no defects were observed.A functional test was performed and the pull force were above the minimum requirements.The manufacturing records were reviewed, and the manufacturing/packaging criteria were met prior to the release of this batch.Per the condition of the samples, no pull off suture needle was found, and the tested samples met the finished goods requirements.
 
Event or Problem Description
It was reported that the patient underwent a dental procedure on an unknown date and suture was used.During the procedure, the needle popped off at the swage.The procedure was completed with an alternate like device with no patient consequences reported.There is no additional information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SILK SUTURE 18"(45CM) 3-0 BLK
Common Device Name
SUTURE, NONABSORBABLE, SILK
Manufacturer (Section D)
ETHICON INC.
p.o. box 151, route 22 west
somerville NJ 08876 0151
MDR Report Key8699754
Report Number2210968-2019-82750
Device Sequence Number19375390
Product Code GAP
UDI-Device Identifier10705031007062
UDI-Public10705031007062
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K946173
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 05/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2023
Device Catalogue Number632G
Device Lot NumberMJH223
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/03/2019
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/20/2019
Initial Report FDA Received Date06/14/2019
Patient Sequence Number1
-
-