| Catalog Number |
TGU454520 |
| Medical Device Problem Codes |
Collapse (1099); Patient-Device Incompatibility (2682)
|
| Health Effect - Clinical Code |
Death (1802)
|
| Date of Event |
05/31/2019
|
|
Type of Reportable Event
|
Death
|
|
Additional Manufacturer Narrative
|
|
The review of the manufacturing paperwork verified that this lot met all pre-release specifications.According to the gore® tag® thoracic endoprosthesis instructions for use (ifu), adverse events that may occur and/or require intervention include, but are not limited to renal failure or other renal complications, and death.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2019 the patient had a thoracic endovascular repair using conformable gore® tag® thoracic endoprostheses.It was reported that the patient had very torturous anatomy and the second device deployed was compressed immediately after deployment.They physician ballooned the device and while the device was slightly compressed due to the patient's anatomy, blood flow was normal.On (b)(6) 2019 the patient presented with renal failure and an aortobifemoral bypass was performed.It was reported that device previously compressed and ballooned had collapsed causing the renal failure.It was reported that the thoracic aorta was s shaped and the lumen size was variable based on how the tortuosity was kinking the aorta.On (b)(6) 2019 the patient expired.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction date of event is (b)(6) 2019.
|
| |
|