Product complaint # (b)(4).Udi: (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Visual examination of the device revealed that the plate had broken at one of the wings.The second half of the broken end of the device was also returned.The fracture analysis report reveals uniform rough morphology of the fractured surface indicating that the instrument underwent a static failure.A review of the device history record was conducted.No issues were identified during the manufacturing and release of this product that could have contributed to the problem reported by the customer.A definitive root cause for the device¿s wing fracturing cannot be positively determined.However, the fracture analysis report reveals uniform rough morphology of the fractured surface indicating that the instrument underwent a static failure due to a single, sudden, unexpectedly high force placed on the wing of the device.As there has been no issue identified in the manufacturing or release of the device that could have contributed to the problem reported by the customer, this complaint file will be closed with no further action required.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|