| Catalog Number |
101204010 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Foreign Body Reaction (1868); Pain (1994); Loss of Range of Motion (2032); Tissue Damage (2104); No Code Available (3191)
|
| Date of Event |
10/18/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Initial reporter occupation: lawyer.(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Pinnacle claim submission form and medical records received.Pinnacle claim form has no allegation provided.After review of medical records on ceramic on metal, the patient was revised due to metallosis and trunnionosis with elevated metal ions.Indications of the procedure stated, the patient developed signs and symptoms and clinical findings consistent with metallosis.Clinical visit on (b)(6) 2016 reported pain, limited range of motion, bursitis, walks with antalgic gait, and metallosis.However, recent laboratory results on (b)(6) 2016 shows cr ion level is below 7ppb.Doi: (b)(6) 2009 - dor: (b)(6) 2017 (left hip).
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary = > no device associated with this report was received for examination.A worldwide complaint database search found no other related reported incidents against the provided product code/lot number combination since release for distribution.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|