| Model Number |
DL-330-10-K |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/21/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).At the time of the event, review of the device ecog and lead impedances confirmed that the ecog data were suggestive of a potential lead break.Investigation of the returned lead confirmed lead damage.The lead coil has a kink about 12 cm from the distal end.Root cause was not identifiable.
|
| |
|
Event or Problem Description
|
|
During a routine review of the patient's neurostimulator ecog data on (b)(6) 2016, it was noticed that there was a signal loss on the left depth lead.The treating clinician stated that the patient's mother indicated that the patient fell from bed in (b)(6).The patient was seen by the treating center on (b)(6) 2016 at which time stimulation was disabled on that lead; however stimulation remained enabled on the other lead and the patient continued to be treated with responsive stimulation.The depth lead was replaced on (b)(6) 2019 during a routine battery replacement.
|
| |
|