| Model Number |
ZKB00 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Blurred Vision (2137); Halo (2227); No Code Available (3191)
|
| Date of Event |
11/05/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported a zkb00 17.5 diopter intraocular lens (iol) was implanted in the patient's left eye on (b)(6) 2017.It was later explanted on (b)(6) 2018 due to the patient experiencing hazing, halos, and glare.The procedure was completed successfully with a non-johnson and johnson lens.There was no incision enlargement or suture required.The patient is happy and doing well post-operatively.No further information was provided.Patient has bilateral implants.This report represents the left eye.The right eye will be submitted under a separate report.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the product was not returned to the manufacturing site; therefore, the complaint issue reported was not verified.Manufacturing records review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specification.Historical data analysis: a search of complaints revealed no similar complaints were received for this production order number.Conclusion: as a result of the investigation there is no indication of a quality product deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
|
| |
|
Search Alerts/Recalls
|