| Catalog Number |
2990001 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/20/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product was not returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient was pre-operatively diagnosed with kyphosis; and underwent l2-l3 posterior lumbar interbody fusion, l4-l5 transforaminal lumbar interbody fusion and l4 pedicle subtraction osteotomy.Intra-op, after hammering the first cage at l4-l5, it was noticed that the claw part of the inserter remained in the cage.This was found in the image after the placement of the cage.The physician tried to remove the broken part by suction of blood.When performing suction, the broken part could not be found in the patient¿s body, and the broken part was found missing.It could not be determined if the broken fragment of the instrument remained inside the patient or not.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis: visual exam and microscopic exam show that the upper tab of the instrument is completely sheared off and the threaded inner shaft is missing.The broken piece was not returned for the analysis.This is consistent with bend stress overload.The hardness of the instrument is within print specification.If information is provided in the future, a supplemental report will be issued.
|
| |
|