Catalog Number W8807
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).A manufacturing record evaluation was performed for the finished device batch mph274-w8807 and no non-conformances were identified.Attempts are being made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.When did the needles pull off the suture threads? in the package/during product removal from package/during use on patient? how was case completed?.
 
Event or Problem Description
It was reported that the patient underwent an unknown procedure on an unknown date and suture was used.The needle is coming off with the minimal pressure making it unusable.There were no adverse patient consequences reported.
 
Additional Manufacturer Narrative
Additional h-3 summary: two empty opened boxes and seventeen unopened samples of product code w8807, lot mph274 were returned for evaluation.The complaint sample was not returned.During the visual inspection of thirteen samples, no defects were found on the packages.The samples were opened, and the swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along of the strands and no defects were observed.A functional test was performed and the pull force were above the minimum requirements.Per the condition of the samples, no performance pull off suture needle was found, and the tested samples met the finished goods requirements.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRLNE BLU 24IN 6-0 D/A CC175-8 MP
Common Device Name
SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE
Manufacturer (Section D)
ETHICON INC.
p.o. box 151, route 22 west
somerville NJ 08876 0151
MDR Report Key8703707
Report Number2210968-2019-82789
Device Sequence Number11280866
Product Code GAW
UDI-Device Identifier10705031231719
UDI-Public10705031231719
Combination Product (Y/N)N
PMA/510(K) Number
K133356
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/21/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2023
Device Catalogue NumberW8807
Device Lot NumberMPH274
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/25/2019
Initial Date Received by Manufacturer 05/21/2019
Supplement Date Received by Manufacturer06/25/2019
Initial Report FDA Received Date06/17/2019
Supplement Report FDA Received Date07/23/2019
Patient Sequence Number1
-
-