Product analysis: analysis was able to confirm the menu dial knob was not working properly.The knob encoder was out of electrical specification.Analysis also noted that the upper case was cracked at the boss, the lower case was damaged, and the display wires were pinched but the insulation was not compromised.All found defective parts were replaced and all the firmware was updated.The device passed all final functional tests.If information is provided in the future, a supplemental report will be issued.
|