| Catalog Number |
M8726 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/01/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The product upon which this medwatch is based has been received, however, the product evaluation is not yet complete.Any further information derived from the evaluation will be submitted in a supplemental 3500a form.The single complaint was reported with multiple events.There are no additional details regarding additional patient events.Attempts were made to obtain the following information.If further details are received at a later date a supplemental medwatch will be sent.Specific number of procedure involved.For each procedure please provide following separately-: event date.Quantity involved.When did each involved device had the needle pulled off? lot involved.The following information was obtained: i was unable to get specific data on cases and dates as this happened over a several week period.Cases were cv and vascular.I was told that the needles were falling off right from the start or very early in the case.Note: events have been reported via mw 2210968-2019-82830, 2210968-2019-82831, 2210968-2019-82832, 2210968-2019-82834, 2210968-2019-82835.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent a vascular or heart procedure on unknown date and suture was used.During the procedure, the needle is popping off very fast.The procedure was completed with another device of the same product code and different lots.There were no patient consequences reported.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Device evaluation summary: twenty-two representative samples of product code m8726, lot # mhp815 were returned for evaluation.The complaint sample was not received.The product code is multistrand count and each packet contains four strands double armed.During the visual inspection of thirteen samples, no defects were found on the packages.The samples were opened, and the swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along of the strand and no defects, or damaged were observed.A functional test was performed and the pull force were above the minimum requirements.Per the conditions of the samples received, no performance pull off suture needle was found, and the tested samples met the finished goods requirements.A manufacturing record evaluation was performed for the finished device lot number mph815, and no non-conformances related to the reported complaint condition were identified.
|
| |
|
Search Alerts/Recalls
|