H3.Evaluation summary: one sample was returned for evaluation contained in its original opened unit pack.Visual examination shows no sign of any defect.The returned unit was inflated and leak tested under water.No leakage detected.The returned unit inflated/deflated without any issue.The returned sample were inflated/deflated three times without any restriction noted.The reported defect could not be detected on the unit returned for evaluation.All of our product under goes a four hour inflate/deflate test and it is at this stage of the process that any defects are removed from the lot.The reported defect could not be detected on the unit returned for evaluation.Information has been added to the database and trends will continue to be monitored.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|