Visual, dimensional, functional and material analysis could not be performed as the device was not returned as the patient wanted to keep the explant so an evaluation was not possible.A review of the device and complaint history records could not be performed as a valid lot number was not provided and could not be obtained.Patient did not perform high level activities post op and did not experience a fall.Per surgical technique: internal fixation appliances are load sharing devices which are used to obtain alignment until normal healing occurs.In the event that healing is delayed, does not occur, or failure to immobilize the delayed/nonunion results, the implant will be subject to excessive and repeated stresses which can eventually cause loosening, bending or fatigue fracture.The degree or success of union, loads produced by weight bearing, and activity levels will, among other conditions, dictate the longevity of the implant.If a nonunion develops or if the implants loosen, bend or break, the device(s) should be revised or removed immediately before serious injury occurs.Root cause of fractured screw that was replaced is most likely due to delayed/non-union.
|